Covedian/Shiley Adult/Pediatric DAR HME (#353U5413) feature a gas sampling port. It is indicated for single patient use on anesthetized and respiratory care patients who require a breathing circuit.
The product is intended to be replaced at least once every 24 hours.
Features
- Hygrolife HME Heat and Moisture Exchanger Hygroscopic
- Hygroscopic HME Media Traps Patient''s Own Moisture, and Minimizes Heat Loss
- Low Air Flow Resistance
- Moisture Output Greater Than 30mg H2O/l (ISO 9360, Vt 750ml)
- Calcium Chloride treated HME Media Improves Effectiveness
- End tidal CO2 Port Offers Convenient Access to Airway Gases
- ISO standard 15 and 22mm fittings
- Supplied clean and individually packaged
Specifications
|
HME
|
Adult/Pediatric HME
|
Adult/Pediatric HME with Extendable Connector
|
Catalog Number (VentureResp.com SKU #)
|
353U19004 (#ML-9004)
|
353U5413 (#SH-353U5413)
|
353U5908 (ML#5908)
|
Quantity/box |
25 |
50 |
25 |
mg H2O/L Moisture Output and Temperature @ 2 hours (ISO 9360) |
Vt 250 ml |
--- |
28.8 |
20.5 |
Vt 500 ml |
[email protected]°C |
[email protected]°C |
31.0 |
Vt 750 ml |
--- |
[email protected]°C |
29.7 |
Vt 1000 ml |
--- |
[email protected]°C |
29.6 |
Resistance to flow (cm H2O) |
at 30 L/min |
0.7 cm H2O |
0.7 cm H2O |
0.5 cm H2O |
at 60 L/min |
1.8 cm H2O |
2.0 cm H2O |
1.1 cm H2O |
at 90 L/min |
3.0 cm H2O |
3.8 cm H2O |
2.0 cm H2O |
Dead space (connectors included ISO 9360) |
16 ml |
36 ml |
60 ml (compressed) 75 ml (extended) |
Weight |
8.5 g |
24 g |
12 g |