The AirLife Edith Trach 1000 Heat Moisture Exchangers (#557005000) meet all the necessary requirements for patients in the OR, ICU and other respiratory care environments, adults and pediatrics. AirLife HME Edith Trach 1000 features a clear casing, making it easy to conclude when the HME needs to be replaced.
Package includes 40 Heat Moisture Exchangers
Features
- Minimal and stable moisture loss over time, helping ensure that patients conserveadequate heat and moisture
- Low breathing resistance, reducing work of breathing for the patient
- Low weight, adding to patient comfort and safety
- Low dead space, helping ensure adequate patient ventilation
- Latex-free materials, offering safer use with all patients
- Transparent housing, helping to determine when to replace the HME
- A wide range of products, meeting various requirements
- Design compliant with ISO standards, standardizing connector sizing
Specification
Technical data |
|
|
|
|
|
Maximum tidal volume (mL) |
500 |
1,000 |
1,500 |
1,500 |
1,000 |
Moisture output (mg H2O/ Measured at VT (mL) |
321 /301 250/500 |
301 /291 750/1,000 |
31.51 /30.51 750/1,000 |
33.511 /32.51 750/1,000 |
242 500 |
Moisture loss (mg H2O/L) Measured at VT (mL) |
5.51 /7.51 250/500 |
7.51/8.51 750/1,000 |
6.01 /7.01 750/1,000 |
4.011 /5.01 750/1,000 |
13.52 500 |
Pressure drop (kPa/cmH2O) At 30 L/min At 60 L/min |
0.08/0.81 0.2/2.01 |
0.1/1.01 0.25/2.51 |
0.1/1.01 0.25/2.51 |
0.05/0.51 0.14/1.41 |
" 0.02/0.2 |
Dead space (mL) |
17 |
28 |
38 |
90 |
16 |
Weight (g) |
6 |
8 |
9 |
20 |
6 |
Connectors Patient side Machine side |
22 M/15 F 15 M |
22 M/15 F 15 M |
22 M/15 F 15 M |
15 F 22 F |
15 F - |
1. Measured according to standard ISO 9360:2000. 2. Moisture loss is determined using dry gas (not room air). These products are in conformity with the relevantprovisions of the Council Directive 93/42/EEC, concerning medical devices, Annex I - Essential Requirements.
Changing frequency:
Change the frequency of Edith HME use after everypatient or every 24 hours, when used continuously onthe same patient.
Contraindications:
Edith HMEs are contraindicated in patients producingfulminating, frothy secretions within their airways andlungs. Edith HMEs should not be used on patients withvery small tidal volumes, such as neonates or very smallchildren. They should not be used with active humidifiersor nebulizers.
Connections:
Edith HMEs intended for artificially ventilated patientsshould be placed between the proximal end of the artificialairway and the Y-piece of the breathing circuit. The EdithTrach should be placed on the tracheostomy tube.